Measuring Gas | Carrier Gas | Basic range | Smallest range | Smallest suppressed zero range | Multi Gas Mode |
H2 | N2 or air | 0% -100% | 0% -0.5% | 98% -100% | Yes |
H2 | Ar | 0% -100% | 0% -0.4% | 99% -100% | Yes |
H2 | He | 20% -100% | 20% -40% | 85% -100% | On request |
H2 | CH4 | 0% -100% | 0% -0.5% | 98% -100% | On request |
H2 | CO2 | 0% -100% | 0% -0.5% | 98% -100% | On request |
He | N2 or air | 0% -100% | 0% -0.8% | 97% -100% | Yes |
He | Ar | 0% -100% | 0% -0.5% | 98% -100% | Yes |
CO2 | N2 or air | 0% -100% | 0% -3% | 96% -100% | Yes |
CO2 | Ar | 0% -60% | 0% -10% | - | Yes |
Ar | N2 or air | 0% -100% | 0% -3% | 96% -100% | Yes |
Ar | O2 | 0% -100% | 0% -3% | 96% -100% | Yes |
Ar | CO2 | 40% -100% | - | 80% -100% | Yes |
CH4 | N2 or air | 0% -100% | 0% -2% | 96% -100% | Yes |
CH4 | Ar | 0% -100% | 0% -1.5% | 97% -100% | Yes |
O2 | N2 | 0% -100% | 0% -15% | 85% -100% | Yes |
O2 | Ar | 0% -100% | 0% -2% | 97% -100% | Yes |
N2 | H2 | 0% -100% | 0% -2% | 99,5% -100% | Yes |
N2 | Ar | 0% -100% | 0% -3% | 97% -100% | Yes |
N2 | CO2 | 0% -100% | 0% -4% | 96% -100% | On request |
NH3 | H2 | 0% -100% | 0% -5% | 95% -100% | On request |
NH3 | N2 | 0% -100% | 0% -10% | 60% -100% | On request |
CO | H2 | 0% -100% | 0% -2% | 99% -100% | On request |
SF6 | N2 or air | 0% -100% | 0% -2% | 96% -100% | On request |
Dimensions without connectors; weight | 145mm x 80mm x 85mm; max. 1800g |
Power supply | 24V DC (18V to 36V), 700mA |
Ambient temperature range | -5°C to 50°C, other on request |
Linearity | < 1% of range |
Warm up time | About 30min; 1h for small ranges |
Flow rate | 40l/h to 150l/h; 60l/h -80l/h recommended |
T90-time | <1sec at flow rate higher 60l/h |
Noise | < 1% of smallest range |
Drift at zero point | < 2% of smallest range per week |
Repeatability | < 1% of range |
Error due to change of ambient temperature | < 1% of smallest range per 10°C |
Error due to change of flow at 80l/h | < 1% of smallest range per 10l/h |
Gas pressure | Max. 2MPa (20bar) |
Error due to change of pressure (above 800hPa abs.) | < 1% of smallest range per 10hPa |